Applications of Class II Biosafety Cabinets
      Type A1 Type A2 Type B1 Type B2
Containment and Protection Provides protection from particulates Personel Yes Yes Yes Yes
Product Yes Yes Yes Yes
Environmental Yes Yes Yes Yes
Provides protection from vapor & gases Personel Non suitable for use with chemicals Only if exhausted to facility exhaust system Reduces exposure Yes
Product Non suitable for use with chemicals No Reduces exposure Yes
Environmental Non suitable for use with chemicals Only if exhausted to facility exhaust system Reduces exposure Yes
Air Flow characteristics Cabinet face velocity Minimum of 75 FPM Minimum of 100 FPM Minimum of 100 FPM Minimum of 100 FPM
Nominal % Recirculated -70% -70% -50% 0%
Exhausted -30% -30& -50% 100%
Plenum Biologically contaminated plenum pressure Negative to room Negative to room Negative to room Negative to room
Exhaust characteristics Cabinet exhaust source Common plenum Common plenum Exhaust plenum Exhaust plenum
To room Yes (no, if vented outside) Yes (no, if vented outside) No No
Vented outside Optional Optional Yes Yes
Connection Type Canopy Canopy Hard ducted Hard ducted

Font: The Baker Company